Surge tester

Per non restare ingarbugliati nella produzione: il surge tester di SPS electronic.

2: Surge- and insulation tester (DC): ST 3810L

Surge test:
• 300 - 6,000 V DC
• Surge capacity: 10, 40, 100, 200 nF
• Partial discharge measurement

Insulation resistance test:
• 100 - 6,000 V DC
• up to 100 GΩ (voltage dependent)
• Safety current limitation

Hipot test DC:
• 100 - 6,000 V DC
• Safety current limitation

4: Switching matrix

Freely configurable for:
• Number of winding connections
• Number of temperature sensors
• Body connection
• Other connections relevant for testing

5: DUT

Customized for your DUT:
• Stator
• Motor
• Rotor

 

What is a surge test?

Surge voltage is a voltage pulse with which the test object is subjected to. A meaningful test is possible on stators, transformers and coils. A test on complete electric motors is only possible on pre-aligned permanent magnet motors, testing on squirrel cage motors is not useful.

What are the advantages compared to hipot testing?

With the surge voltage test the windings can be tested among each other and the windings against ground / housing. Insulation faults within the windings can also be detected. With a mere hipot test, the windings are tested against earth / housing, but not the windings among each other.

How is the surge voltage generated?

After connecting a charged capacitor with high voltage, the energy stored in the capacitor discharges into the inductance of the test object winding and back into the capacitor. This results in a sinusoidal curve, which is produced by the damped oscillation. Each inductance to be tested has a specific frequency and a characteristic amplitude curve.

How is it measured?

The surge test is currently carried out on our products at a voltage of max. 6 kV. In practice, there are two methods of measurement: the master curve method and the double pulse method:

• Double impulse:

Used for 1-phase coils / windings. Two pulses are applied to the DUT one after the other and these are compared. Flashovers and insulation faults are detected by means of the two curves. Incorrect numbers of windings cannot be detected.

• Master curve:

Recording of a master curve with a sample DUT. Incorrect numbers of windings are detected and a comparison of several windings (e.g. with a 3-phase DUT) is possible. Flashovers and insulation faults are detected by means of the curves.

How can the result of the test be evaluated?

The evaluation methods "Defect area", "Differential defect area" and "Tolerance band" are available. Evaluations are possible both with the "master curve method" and with the "double pulse comparison method".

Figure 1: wrong number of windings

If the DUT has a lower number of turns than the sample DUT, the oscillating circuit has a higher frequency. If there are too many windings, the opposite is true.
Figure 2: Short circuit between turns

The error pattern in the case of a short circuit of the windings is similar to that of a too low number of windings, but the curve is much more attenuated.
Figure 3: Wire breakage

When a wire breaks, the curve oscillates similar to the master curve, but is more damped and then stops abruptly.
Figure 4: Insulation fault

Missing insulation from the housing

Prodotti su misura

Come richiesto

Sistemi di test personalizzati in base alle esigenze personali.

Per saperne di più

All-in-one!

Tester multifunzione

Desideri eseguire test multipli con un unico dispositivo

Vai ai tester multifunzione

Contatti

Ti serve aiuto? Saremo lieti di offrirti la nostra consulenza!

Sono interessato a: