4.3‘‘ colour TFT display: 1800 series
Modern digital technology and safety current limitation for unproblematic workspace designs.
Services
Customized
Pre-owned devices
Company
The new multifunction safety testers of the 1805 and 1886 series combine all the functions of their predecessors and expand them with many features that make your work much easier. These include, for example, the new touch display. This allows all functions to be operated and the results to be read directly at the same time. Thanks to the interfaces available ex works, the 1805 and 1886 series can be effortlessly integrated into your production line or operated "standalone".
Hipot Safety Analyser LG 1805B ![]() | Hipot Safety Analyser KT 1886B ![]() | Hipot Safety Analyser KT 1886J ![]() | |
Type of device | Desktop device | Plug-in 19‘‘ / 5 HU | |
High voltage AC | |||
Voltage range | 100 – 5,500 V | ||
Current range | 0 – 3 mA | 0 – 100 mA | |
High voltage DC | |||
Voltage range | 100 – 6,000 V | ||
Current range | 0 – 10 mA | 0 – 100 mA | |
Insulation DC | |||
Voltage range | 100 – 6,000 V | ||
Measuring range | 250 kΩ – 1 GΩ | ||
ARC Detection (optional) | 0 – 100% | ||
Ground bond | 1 – 30 A / 0 – 10 Ω | ||
Equivalent leakage current | 10 – 270 V / 0 – 10 mA | ||
Continuity | 24 V DC / 0 – 600 mA | ||
Resistance | 24 V DC / 1 – 1,000 Ω | ||
Source function test | |||
Voltage | - | 50 – 270 V | |
Frequency | - | 50 Hz and 60 Hz | |
Power / current | - | 250 VA / 1 A | 500 VA / 2 A |
Function Test | |||
Current AC | 0 – 16 A | ||
Current DC | 0 – 16 A | ||
Voltage AC | 0 – 300 V | ||
Voltage DC | 0 – 300 V | ||
Active power | 0 – 4,000 W | ||
Reactive power | 0 – 4,000 VAR | ||
Apparent power | 0 – 4,000 VA | ||
Cos φ | -1 to +1 | ||
Hipot Safety Analyser KT 1885K ![]() | |
Type of device | Plug-in |
High voltage AC | |
Voltage range | 100 – 5,500 V |
Current range | 0 – 3 mA |
High voltage DC | |
Voltage range | 100 – 6,000 V |
Current range | 0 – 10 mA |
Insulation DC | |
Voltage range | 100 – 6,000 V |
Measuring range | 250 kΩ – 1 GΩ |
ARC Detection (optional) | 0 – 100% |
Ground bond | |
Measuring range | 0 – 10 Ω |
Test current | 1 – 30 A |
Equivalent leakage current | 10 – 270 V / 0 – 10 mA |
Continuity | 24 V DC / 0 – 600 mA |