Any medical products which have to be tested for electrical safety in accordance with IEC 601 can be connected to this system. The main tests are:
• earth bond test
• insulation resistance tests (line side)
• insulation resistance tests (at an application part/element of DUT)
• high voltage tests (line side)
• high voltage tests (at an application part/element of DUT)
• line leakage tests (according to IEC 601)
With line leakage tests you can perform earth, housing and patient line leakage measurements. With patient line leakage measurements you can also perform measurements with power supply voltage at the application part or element of the DUT. In addition the simulations (N.C. / S.F.C.) required by standard for the products B, BF or CF can be switched. The voltage range for the line leakage tests can be adjusted via a programmable source in the range of 50 up to 270 VAC.